Insulin Over-Delivery Risk Prompts Medtronic to Issue Global Recall

☞ | View the text of this post. Exclusively available to ASQ Members.

Medtronic is recalling specific lots of MiniMed infusion sets after learning of the risk of insulin overdelivery. The global, voluntary recall affects older insulin pump infusion sets that featured a vent membrane susceptible to blockage and consequently to overdelivery of insulin.

Patient field reports and a root cause analysis suggest overdelivery of insulin occurs shortly after an infusion set change if the vent membrane becomes blocked during the priming process. Medtronic thinks such blockages happen when